62
Maintenance / Further information
G-30-1439-en
OPMI® VISU 200 BrightFlex™ Surgical Microscope
Issue 3.0
Printed on 21. 11. 2002
Trouble-shooting table
•
This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety
must be checked once every 12 months. We recommend having this
check performed by our service representative as part of regular main-
tenance work.
If a failure occurs which you cannot correct using the trouble-shooting
table, attach a sign to the instrument stating it is out of order and con-
tact our service representative.
VISU 200 BrightFlex™ Surgical Microscope
Problem
Possible cause
See
No function at all.
Power cord of suspension
system not plugged in.
Plug in power cord.
-
Power switch of suspension
system not on.
Press power switch. Green
light in power switch must
come on.
-
Circuit breaker in suspen-
sion system power switch
activated.
Press power switch again.
-
Line power failure.
Contact
in-house
electri-
cian.
-
Lamp module has no con-
tact.
Insert lamp module as far as
it will go.
-
Surgical field illumination at
microscope inoperative.
Illumination system of sus-
pension
system
not
switched on.
Switch
on
illumination
system
of
suspension
system.
-
S light guide not properly in-
serted in lamp or micro-
scope.
Insert S light guide as far as
it will go.
-
Failure
of
suspension
system electronics.
Illuminate
surgical
field
using an OR illuminator.
Contact service dept.
-