42
Description
G-30-1439-en
OPMI® VISU 200 BrightFlex™ Surgical Microscope
Issue 3.0
Printed on 21. 11. 2002
Tube and eyepieces for assistant's microscope
7
Push-in widefield eyepiece
Note:
When you remove the eyepiece from the tube, please note that this
eyepiece is fitted with a magnetic coupling. When mounted, the eye-
pieces display a very weak magnetic field. For this reason, the usual
rules for the handling of magnets must only be observed with eye-
pieces which have not been mounted on the microscope:
•
Do not place the eyepieces close to instruments where there is any
risk of magnetization.
•
Do not place eyepieces on sensitive electronic instruments such as in-
fusion pumps, cardiac pace-makers, measuring instruments or mag-
netic data carriers such as disks, audiotapes and videotapes or credit
cards.
•
Always store the eyepieces in their original packaging when not in
use.
8
Eyecup
Always adjust the eyecups in such a way that you can see the full field
of view.
9
Diopter scale
For reading the prescription set.
10
Diopter setting ring
The eyepieces provide ametropia compensation from -8 D to +5 D.If
you wear your eyeglasses during work, adjust the ring to 0 D. Turn the
ring until you have found the optimum setting. A built-in catch lever
locks the ring in the position set.
11
Binocular assistant's microscope
Note:
Always use the assistant's microscope with the appropriate (shown)
binocular assistant's tube. If you use a different tube, you will obtain
an inverted image.
–
Viewing with eyeglass-
es:
Screw in eyecups all the way.
–
Viewing w/out eyeglass-
es:
Screw eyecups outward until you see
the full field of view.