Iron Man User Guide v. 1.4 

 

 

6)

 

Adjusting PMTs  

 

These guidelines assume a fixed sample.  For live samples, you should try to avoid Auto Expose or 
Continuous modes as these will expose your sample to large amounts of excitation light prior to image 
acquisition.  Instead use Live mode or optimize in one region of slide 
and move to another for data acquisition. 

 

This step depends on your experimental design and question.  To 
ensure that your images are quantitative, you must avoid over- or 
understaurating pixels, as you will not be able to properly measure 
intensity in this case.  It is also important to remember that your 
brightest sample may be a particular Z plane, later slide, alternate 
timepoint, and these may be oversaturated when you get to them 
later.   
-

 

Select the first channel to optimize (check on/off in Imaging 
Setup area).   I usually prefer to optimize one channel/track 
at a time, but if you plan to use multiple lasers on a single 
track, make sure these will all be on during optimization. 

-

 

Ensure that your sample is on the brightest Z plane – go to 

live mode

 

and adjust using focus knob.   

-

 

Set lasers to appropriate starting values

 

- for the Argon 

and Diode lasers, 3-5% is usually sufficient, and for the 
HeNe lasers, 25-35% is usually a good starting point. 

-

 

Set the pinhole size to 1AU

.  Unless you have a deliberate 

reason to change this, 1AU is good for most experiments.  
(Bigger pinhole = more light = less “confocal”).  Don’t go 
smaller.  Bigger is useful for if you need to image a really 
thick sample and do not want to image many 1AU confocal 
slices.   

-

 

Click 

Auto Expose

.  The software will now attempt to set 

the gain and offset for that channel.  

-

 

Click 

Continuous

, and select the 

Range indicator Look Up 

table

 beneath the image.  Oversaturated pixels will appear 

red and undersaturated pixels will appear blue.  You can 
manually adjust the gain and offset to achieve a good 
balance between over- and undersaturation for your 
experiment.  You should adjust the 

offset

 

first.  This will 

adjust the background.  Be aware that if you change this too much you 
will also remove signal form your sample.  A little goes a LONG way, go 
down to add undersaturated pixels, go up if you have too many 
undersaturated pixels.  After adjusting the offset, adjust the 

gain 

(master)

.  This will adjust the foreground and background signals.  If you 

are at the high end of gain, bump up laser power; if you are at low end, 
reduce laser power.  

-

 

Repeat these steps for all channels in your experiment.  

Summary of Contents for LSM 510 Inverted

Page 1: ...EC Plan Neofluar 20x 0 8NA Plan Apo Chromat 40x 1 3NA oil DIC EC Plan Neofluar 63x 1 4NA oil DIC Plan Apo Chromat 100x 1 4NA oil DIC Plan Apo Chromat Standard coverslip thickness 0 17 1 5 coverslip Me...

Page 2: ...Tab 6 At the top In the Setup Manager window In the Imaging Setup window In the Channels window In the Online Acquisition window Adjusting PMTs Snap Z series The Information tab next to the image 11 W...

Page 3: ...ment 8 Note the details in Information tab next to your image these are useful to write in your imaging notes 9 Save and copy your data The LMCF is not responsible for long term data storage See Appen...

Page 4: ...including lasers 2 Mercury Arc lamp located underneath floating air table 3 PC 4 The microscope green switch on right side of microscope base If you only need to access the PC only turn on 1a and 3 L...

Page 5: ...u are not looking through the eyepieces To look at your slide 1 Load the slide coverslip down onto the slide holders on the stage these can be adjusted manually Please be sure that your slide is clean...

Page 6: ...when it is done Choose On Open the Laser Properties window by clicking on the arrow and adjust laser output to 40 Do not go over 40 output or over 6 1 A Do not keep changing this Output set it for yo...

Page 7: ...k will be faster but may lead to bleedthrough of your signal across channels An example of both a single track multi channel acquisition and multi track acquisition are shown To add or remove tracks c...

Page 8: ...ges and can allow you to reduce your scan speed Bit Depth the number of grey levels the image will contain Higher bit depth will result in greater signal dynamic range greater discrimination between p...

Page 9: ...ing values for the Argon and Diode lasers 3 5 is usually sufficient and for the HeNe lasers 25 35 is usually a good starting point Set the pinhole size to 1AU Unless you have a deliberate reason to ch...

Page 10: ...ion until desired distance is reached click set last o Repeat in the other direction click set first o Either choose number of slices or step size The software will suggest the optimal step size This...

Page 11: ...objectives you may have used with lens paper If you are at all unsure of this process ask for help 3 Leave the microscope on a low power objective for next user If someone is signed up within the next...

Page 12: ...e and password 5 Click Ok and the drive should now appear Copy over your files If you have a lot of files be sure to allow time for the transfer Under ideal conditions you will be able to copy close t...

Page 13: ...Iron Man User Guide v 1 4 13 APPENDIX B Parts of the microscope not all these components are present on our microscope...

Page 14: ...o bleedthrough of signal You can do a combination of multi track and multi channel for example image DAPI and red together in a mult channel track and image green separately in its own track Load a pr...

Page 15: ...Resources tab at the bottom Life Technologies SpectraViewer is a straighforward and reasonably complete one Excitation spectra are shown as dotted lines and emission spectra are solid Next figure out...

Page 16: ...transmit the emission from the sample through 405 488 405 514 594 405 488 543 633 458 458 514 488 543 488 594 NFT 80 20 neutral density filter passes 80 and blocks 20 2 Sample 3 Dichroic split emissio...

Page 17: ...l reflect all light below 543nm to the right and allow to pass through all wavelengths greater than 543 The following plot illustrates the emission spectra of our three fluorophores with a green line...

Page 18: ...nels It is important to note the overlap in emission spectra In this example it is possible that some of the DAPI emission will show up in the FITC channel and some FITC emission will show up in the R...

Reviews: