Change and adjustment of the 6 V / 10 W halogen lamp.
•
Place the microscope on its side on the table.
•
Loosen the knurled screw (4-2/
1
) and pull out
the illuminator (4-2/
4
).
•
Remove the halogen lamp (4-3/
1
) from the
metal clamps (4-3/
2
).
•
Use the plastic protection bag supplied by the
manufacturer to hold the lamp and insert the
lamp in the metal clamps (4-3/
2
). Remove the
fingerprints before switching on the lamp.
Adjusting the halogen lamp:
•
For an exact adjustment, the lamp coil and its
reflection image must be parallel to each other
and be identical in size when looking on the
lamp and concave mirror. To achieve this,
loosen screw (4-2/
3
) and move halogen lamp
with mount (4-3/
3
) until the required setting
has been obtained.
•
Retighten screw (4-2/
3
).
•
Then turn screw (4-3/
4
) until the coil and its
reflection image appear identical in size (Fig.
4-3/
6
). After adjustment, insert the lamp
housing and tighten the knurled screw (4-2/
1
).
Fig. 4-2
Changing the halogen lamp
Fig.4-3
Adjusting the lamp coil
Summary of Contents for KF 2 ICS
Page 1: ...KF 2 ICS Transmitted Light Microscope Operating Manual...
Page 6: ...Carl Zeiss KF 2 ICS VI B 40 028 e 10 97 Overall view of the KF 2 ICS microscope...
Page 8: ...Carl Zeiss KF 2 ICS 1 2 B 40 028 e 10 97...
Page 10: ...Carl Zeiss KF 2 ICS 1 4 B 40 028 e 10 97 1 3 Modules Fig 1 1 KF 2 ICS microscope equipment...
Page 14: ...Carl Zeiss KF 2 ICS 2 2 B 40 028 e 10 97...
Page 20: ...Carl Zeiss KF 2 ICS 3 2 B 40 028 e 10 97...
Page 32: ...Carl Zeiss KF 2 ICS 4 2 B 40 028 e 10 97...