Safety notes
Instructions for Use
Risk of injury from rolling of the floor stand! Always push rather than pull the microscope to
move it. Lock the casters during operation.
Always switch off the system before disconnecting it from / connecting it to line power, for
cleaning its surface or if it will not be used for a prolonged period of time;
Before an extended period of non-use, remove the battery from the remote control;
Avoid looking directly into the light source, e.g. into the microscope objective lens or into the
light guide, in order to prevent damage to the eye. Do not shine lights in anyone's eyes.
EXTARO 300 can only be used in professional healthcare facility environments such as physician
offices, dental offices, clinics, hospitals (emergency rooms, patient rooms, intensive care and
surgical rooms without high-frequency surgical equipment). Additionally, it should be used
outside the RF shielded room of medical electrical systems for MRI.
Use of EXTARO 300 adjacent to or stacked with other equipment may result in improper
operation. If such use case cannot be avoided, the user must verify that the EXTARO 300 and
its adjacent equipment are functioning normally before operation.
Depending on the type of application, take appropriate precautions to ensure that the surgical
procedure or treatment can still be completed even if the full functionality of this microscope
(e.g. illumination) is not available.