Instructions for Use
Trouble shooting
Fault
Cause
Rectification
Camera installed but no
image output on the
monitor
HDMI connector not in well
contact with the HDMI
socket.
Plug the connector in the HDMI socket
correctly.
Defective CCU
Contact Service.
No video image, only
color bars
Camera head not connected
Connect the camera head
Image saving failed
No USB data carrier inserted. Insert USB data carrier.
USB data carrier is inserted
but there is insufficient
memory space.
Delete some files on the USB data
carrier to release some memory space.
USB data carrier does not
contain any file system or a
damaged file system.
Format the USB data carrier or
replaced it.
Failure of the Mode Control knob
In the event that switching between illumination modes by clicking the Mode Control knob doesn’t
take any effect:
1)
Power off the system.
2)
If user still wants to continue work with the OPMI before fixing the issue, he/she can use the
Emergency knob to return to normal white light mode (e.g. use a coin to rotate it), and then
power on the system again.
3)
Contact Zeiss Service.
Failure of the motorized filter wheel (optional)
In case the motor of the filter wheel assembly cannot move:
1)
If user still wants to continue work with the OPMI before fixing the issue, he/she can use the
Emergency knob to rotate the filter wheel (e.g. using a coin).
2)
Contact Zeiss Service.
305935-3500-000 – 2.1 – 2017-12-15
42 / 59