Instructions for Use
Technical specification
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic
environment – guidelines
Voltage dips, short
interruptions and
voltage variations as
per IEC 61000-4-11
0% U
T
; 0.5 T (0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°)
0% U
T
; 1 T (0°)
70% U
T
; 20 T (0°)
0% U
T
; 0.5 T (0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°)
0% U
T
; 1 T (0°)
70% U
T
; 20 T (0°)
The quality of the supply
voltage should be that of a
typical business or hospital
environment. If the user of
EXTARO 300 requires
continued function even in the
event of interruptions in the
power supply, we recommend
to power the device EXTARO
300 from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz)
Magnetic field as per
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
typical of business and hospital
environments.
NOTE: UT is the AC supply voltage prior to application of the test level.
11.8.3
Electromagnetic immunity for non-life-supporting ME equipment and ME systems
The device EXTARO 300 is intended to be operated in one of the electromagnetic environments
specified below. The customer or the user of the device EXTARO 300 is responsible for ensuring that it
is operated in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
Guidelines
Portable and mobile radio communication
equipment should not be used closer to the
device EXTARO 300, including cables, than
the recommended safety distance that is
calculated using the equation applicable to
the transmission frequency involved.
Recommended safety distance:
Conducted RF
disturbances as
described in IEC
61000-4-6
3 V
RMS
150 kHz to 80
MHz
3V
d = 1
.2
Radiated RF
disturbances as
described in IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1
.2 for 80 MHz to 800 MHz
d = 2
.3
for 800 MHz to 2.5 GHz
305935-3500-000 – 2.1 – 2017-12-15