![Zeiss Crossbeam 350 Instruction Manual Download Page 27](http://html.mh-extra.com/html/zeiss/crossbeam-350/crossbeam-350_instruction-manual_927679027.webp)
ZEISS
3 Product and Functional Description | 3.2 Main Components
5
Bypass
6
Specimen chamber
7
Vent valve, flanged to TIV (VP mode option)
8
Turbo isolation valve (TIV, VP mode option)
9
Turbo pump
10
Ion getter pumps (IGPs)
11
Needle valve (VP mode option)
12
Pirani gauge (VP mode option)
13
Penning gauge
14
Vent valve
15
Roughing backing valve (RBV)
16
Pre-vacuum pump
The microscope can be operated in different vacuum modes:
§
High vacuum (HV) mode
§
Variable pressure (VP) mode, optional
HV mode characteristics
Pressure regions
HV mode is the standard mode of the micro-
scope. HV mode offers the possibility of analyz-
ing and mapping conducting specimens. In HV
mode, the pressure in the specimen chamber is
less than 10
-6
mbar.
Instruction Manual ZEISS Crossbeam 350 | en-US | Rev. 3 | 349500-8111-000
27