5 Operation
Instructions for Use
CLARUS 500
5.5 Acquire Images
78 / 144
2660021165005 Rev. A 2017-08
2. Install the external fixation target (1) onto mount (2) on the pa-
tient support (3).
2
1
3
3. Position the target where you want the patient to focus.
5.5.11 Checking Scan Quality
Once data acquisition is complete, scans must be reviewed for ac-
ceptance prior to being released for analysis.
Reivew all scans carefully. If possible, retake scans that do not pass
the acceptance criteria (Deleting or Replacing an Image [
).
Action
1. Check that the focus is sharp and clear.
2. Ensure that the image is illuminated uniformly.
3. Ensure that there are few to no artifacts that cast shadows on
the image.
5.5.12 Deleting or Replacing an Image
Prerequisite
þ
You reached the image check step in a
Scan Acquisition
task.
þ
You want to delete or replace an image.