5 Connecting the Camera | 5.6 Function indicator
ZEISS
2. Plug the connector
Trigger (1)
in the socket
Trigger Out
of the SVB (Signal distribution
box).
Fig. 6: Signal distribution box (SVB) backside
3. Plug the connector
GPO 0 (2)
in the socket
Trigger In 1
of the SVB (Signal distribution
box).
4. Plug the connector
GPO 1 (3)
in the socket
Trigger In 2
of the SVB.
5. Connect the connectors
GPOPWR (5)
and
AUXPWR (6)
of the trigger cable together.
à
GPO 2 (4) is not connected.
5.6 Function indicator
Signal color
Description
Blue
USB 3.0 connection (Camera) connected with USB 3.0 connection (PC)
Power supply (USB 2.0 cable) attached
Recommended configuration for best camera performance
Green
USB 3.0 connection (Camera) connected with USB 2.0 connection (PC)
Power supply (USB 2.0 cable) attached
Violet
USB 3.0 connection (Camera) connected with USB 3.0 connection (PC)
Power supply (USB 2.0 cable) not attached
Yellow
USB 3.0 connection (Camera) connected with USB 2.0 connection (PC)
Power supply (USB 2.0 cable) not attached
Red
Power supply attached, no driver loaded, camera not yet initialized
No light
No power supply or software shut down (camera in power down
mode)