5 Connecting the Camera | 5.3 Mounting the camera on the microsope
ZEISS
4. Connect the appropriate power connector cable to one of the open connectors from your
computer’s power supply.
5. Connect the other end of the power cable to the PCIe interface card.
6. Insert the PCIe interface card into the appropriate slot on your computer.
7. Close the computer case and reconnect the peripherals.
You can now mount the camera on the microscope and connect it to the PC via the USB 3.0 in-
terface, see chapter
Connecting the camera to the PC [
.
5.3 Mounting the camera on the microsope
To mount the camera onto your microscope’s TV port, use a C-mount adapter. The adapter is not
included in the shipment. You will find some suitbale examples for adapters in the list below:
Port
Name
Order Number
44
Video Adapter 44 C 2/3" 1,0x
452995-0000-000
60
Camera Adapter 60 C 1" 1,0x
456105-9901-000
60
Video Adapter 60 C 2/3" 0,63x
000000-1069-414
60N
Camera Adapter 60N-C 1" 1,0x
426114-0000-000
60N
Camera Adapter 60N-C 2/3" 0,63x
426113-0000-000
NOTICE
Loss of warranty
The Axiocam 512 color is delivered with an integrated IR filter (infrared cut filter). The Axiocam
512 mono is delivered with an integrated BK7 protective glas. This will protect the camera
against dust and reduces optical interferences.
4
The IR filter is situated approx. 5 mm behind the outer edge of the C-mount opening in
the camera.
4
Due to this pre-mounted IR barrier filter, C-mount adapters that screw more than 5 mm
into the thread cannot be screwed into the camera.
4
Do not remove the filter or the protective glas otherwise the warranty will be lost.
1. Remove the dust cap from the camera’s C-mount port.
2. Screw the adapter in as far as it will go.
3. Mount the camera onto the microscope’s TV port. Ensure that no dust enters the opening
of the camera or the microscope’s TV port.
5.4 Connecting the camera to the PC
1. Lay the USB cable (2 in 1) carefully between the camera and the PC.
2. Connect the USB cables to the USB 3.0 interface card of the PC.
3. Connect the USB cables with the camera.
4. If you switch on the PC, the camera will be also switched on.
Power is supplied via the USB 3.0 cable. The USB 2.0 cable supplies the peletier cooling. By
switching off the PC, the camera will be switched of as well. If the camera is installed cor-
rectly, the status LED will be red. If you start the software the LED will turn blue. You`ll find
a detailed description of the LED status light in the chapter
.
User Guide ZEISS Axiocam 512 color/mono | en-US | Rev.1
16
Summary of Contents for Axiocam 512 color
Page 1: ...User Guide ZEISS Axiocam 512 color mono ...
Page 23: ......