zebris Medical GmbH
FDM Technical Data and Operating Instructions
Page 19/54
3.6
zebris SYNC
The
zebris SYNC
serves as a standard solution for the synchronisation of the FDM system
with measuring systems of other manufacturers.
The
SYNC-IN
and
SYNC-OUT
socke
ts provide inputs and outputs for support of „frame by
frame“ In- and Out synchronization. Both sockets are galvanic protected from the platform.
WARNING
Patient’s safety is guaranteed by means of galvanic separation acc.
to to the provisions of IEC 601-1 when a third party device is synchro-
nized with the FDM system. This allows non medical equipment to be
synchronized with the FDM system as long as such devices are out
of patients reach. Nevertheless the user is completely responsible for
the safety of all third party devices used in combination with the FDM
system.
The correct synchronisation of the measuring data of all coupled sys-
tems must be verified before evaluation as soon as a coupling with
devices is carried out, that have not been manufactured by zebris.
zebris does not assume any warranty for the correct functioning and
reliability of the system if the clock signals of external devices do not
correspond to the indicated specifications.