7
Rear Panel
1.
Foot Switch Connector
: For starting a test. Any closed contact switch may be used to start a test by
shorting pins 1 & 3
Mating Conn. & Pins
–
Tyco Electronics 1-480305-0 (
2.
Serial Output
: Connection for interfacing Isaac to computers or printers. (
3.
Option Space
: This space is used for Clamping (pneumatic outputs) or I/O connection
(shown)
If the I/O space is filled clamps will be in the secondary air connection location.
4.
Pressure Regulator
: Controls the air pressure applied to the product under test.
5.
Power Entry Module
: This component includes the power cord socket, on/off switch, and the fuse
holder. Power entry is capable of a range of 109-255~VAC. 2A Fuse (
6.
Secondary Air Connections
:
These positions have multiple uses; Flow control for ramping burst/crack
pressure, coupling or pilot air entry, or secondary test air supply. See the Quick Start shipped with the
tester for details.
7.
Supply
:
(Air or other test gas) Connection is 1/8
”
-27 F NPT (Similar to R1/8 BSPT British Standard Pipe
Thread) Supply gas must be clean and dry.
8.
Ethernet Connector
:
ASCII communication, data output and command string input.
Additional items may be installed on the rear panel, see the quick start guide
shipped with the tester for additional details.
1
2
3
4
5
6
6
7
8