21
Units
There are six selectable engineering units on the setup screen.
These units are:
o
psig
–
pounds per square inch gauge.
o
mbar
–
millibars.
o
mmHg
–
millimeters of mercury.
o
inH2O
–
inches of water column.
o
kPa -
kilopascals.
o
cmH20
–
centimeters of water
o
inHg
–
inches of mercury
The atmospheric pressure used in the
‘sccm’
leak rate calculation. The default value is set to the
pressure at sea level 14.70psia.
The Electronic Regulator response: 10000 (default) a smaller value makes the regulator react quicker in
its feedback state, a larger value slows the response.
Test Pressure and Result Pressure resolution can be modified by selecting the appropriate radio button.