14
Setup Screens
Program
Setting Test Pressure
To set the test pressure, press the data field and enter the test pressure value on the keypad.
For machines with the electronic regulator option see below for test pressure setting instructions.
To Verify Test Pressure
To verify the manual regulator is set correctly, cap the test port and check the
‘Valve On’
box
above the appropriate channel. The current pressure will be displayed in the box.
Units with the electronic regulator will automatically adjust during the test.
Pressure Tolerance ±
The amount of tolerance both plus and minus can be entered in the data fields to the right of the
test pressure. This is the amount of tolerance on the test pressure. During the Fill step a value
outside the boundaries will result in a Hi or LO pressure error. During the Settle step, any
pressure value outside of the boundary will result in an error.
Test Channel Enable/Disable
To enable/disable channels (multi-channel models) check boxes enable/disable channels to be
tested. Sequential models will only allow one selection.
Next Program (Linking Tests)
To link programs together enter the number of the next program to jump to upon a pass result of
the current test. Entering same number as the current test will not loop the tests. A test failure
will cancel the jump to the next program.
The darkened button
shows the active
selection.
Press the ‘i’ to get
more information
about the settings