27
Electronic Regulator
(Optional Item)
With the optional electronic regulator installed adjusting pressure values is done through
the software via buttons on the touch screen. The calibration procedure will follow as listed in
this manual; the only difference is how the regulator is adjusted to achieve the set points.
In the calibration screen and additional button is available:
This button
will open the regulator control screen.
Six buttons will be used for increasing or decreasing the output pressure of the regulator.
+ Small Increase, ++Medium Increase, +++ Large Increase
- Small Decrease, -- Medium Decrease, ---Large Decrease
The buttons on the left side (0 through 10) are the previous settings of the regulator for each
numbered calibration point; these buttons can be
used to ‘jump’
to the previously set pressure
before beginning the fine tuning process.
A live pressure reading is shown in the upper right corner.
Once the de
sired pressure is maintained pressing the ‘
Back
’ button will return to the c
alibration
screen where Cal Points can be set.
Adjustment buttons