18
Pre-Test
Op1
Pre-Test
Op2
Pre-Test
Op1
Pre-Test
Op2
Leak
Test
Cycle
Initiated
Cycle
End
Clamp
Valve 1
Clamp
Valve 2
Status During Test
: Selecting the fixture valves to be off during test will turn off the output during the
‘TEST’ step. The default action is ‘ON’
The typical use is on an occlusion (blockage) test type where the clamp valve will be used to release the
part to allow pressure drop to occur to meet test limits. (Test details)
Hold clamps on Fail
: When this option is selected the clamps will remain in their test position, the fail
indicator LED will flash to alert the operator of the failure. The operator will acknowledge the failure by
pressing the fail LED and the clamps will then follow the set release timer.
Note: The Pass/Fail results sent out the I/O will not be posted until the failure has been acknowledged.