Appendix 1 - Unpacking and Inspection
Inspect the outside of the shipping carton for damage. If damage is detected, contact
the carrier immediately. Remove the 5X00 from the shipping container. Be care-
ful not to discard any parts or supplies. Confirm that all items on the packing list
are present. Inspect all assemblies and components for damage. Save the original
packing carton. Carriers typically require proof of damage due to mishandling.
The following components are included with the purchase of the 5X00 instrument:
5X00 instrument
006515 Flange Mounting Kit
006506 Desiccant Kit
605877 Instruction Manual
605230 Compression Plugs, 2 each
605229 Compression Plugs, 2 each
605804 Compression Plugs, 4 each
605898 Compression Plugs, 3 each
655384 Dual Male RJ-45 Cable Assembly
655383 Adapter, DB-9 to RJ-45
605880 AC Power Cord, 155 VAC U.S. Plug (AC version only)
605944UL Pluggable Socket Connector for AC Power, 3 pole (AC version only)
506939UL Pluggable Socket Connector for 12 VDC Power, 2 pole
605941UL Pluggable Socket Connector for RS-485, 5 pole
605943UL Pluggable Socket Connector for relays, 12 pole, 2 each
605940UL Pluggable Socket Connector for Auxiliary Inputs, 3 Pole
605942UL Pluggable Socket Connector for DO and AUX inputs, 8 Pole, 2 each
If any parts are damaged or missing, contact YSI or your YSI representative im-
mediately.
Unpacking and Inspection
Troubleshooting
Appendix 2 - Troubleshooting
Sensor Troubleshooting
-
When troubleshooting sensors, disable the control/alarm relays.
-
Verify 5X00 wiring and configuration.
Issue
possible cause
possible solution
inaccurate or fluctuating
sensor values
improper placement of probe
assembly
See
Location Considerations
- Step 2 Locate and Install
Probe Assembly
fouled membrane or sensor
cap; damaged membrane or
sensor cap
visual inspection of the
probe assembly required.
Clean membrane or sensor
cap.
improper placement of 5X00
See
Location considerations
- Step 1 Install 5X00 -
bad sensor
bad DO membrane
recalibrate, change mem-
brane, clean sensor, test
sensor outside system
stream in solution of
known value, replace sensor
sensor out of calibration
See
Calibration
DO elevation configuration
incorrect
Correct DO elevation at
Menus → System → Sensor
Setup→DO→Elevation
menu
DO values are affected by
temperature. If DO values
are incorrect, verify that
the temperature source is
accurate and reporting cor-
rect values or that the user
defined temperature value
is correct for the DO sensor
location. If the temperature
sensor is reporting incorrect
values, the DO mg/L values
are not reliable.
See
DO and temperature
-
DO-galvanic or polaro-
graphic inaccurate values
DO sensor wires are termi-
nated incorrectly
see
Wire DO sensors