220
Issue
possible cause
possible solution
probe assembly ground wire
or temp sensor wires (if not
in use) are causing interfer-
ence)
cut back unused wires to
cable sheathing to avoid
shorting.
“over/under” sensor
readings
bad sensor
bad membrane
clean sensor, replace
membrane, calibrate sensor,
test sensor outside system
stream in solution of
known value,
replace sensor
no sensor value on
screen
sensor not enabled
See
Sensor Setup
display setting configuration
set for multiple Run Screens
See
Display Setting
“check probe” message at
sensor calibration
improper placement of probe
assembly, incorrect sensor
wire connection(s), improper
placement of 5X00, bad sen-
sor, and/or bad membrane
See
Location Considerations
Probe Assembly -
See Wire sensors - 49.
See
Location Consider-
ations 5X00 - 27
clean sensor, calibrate,
test sensor outside system
stream in solution of
known value, replace mem-
brane and/or replace sensor
if necessary
Analog aux value is
overlapping to adjoining
sensor window
1X1 & 2X2 display format
setting cannot display values
over 7 digits in aux sensor
window
Adjust aux display format
to lower resolution i.e. from
0.00 format to 0 format.
See also
Notes: Display
Screen Limitations
Relay (Outputs) Troubleshooting
-
Disable control, alarm, and timer relays when troubleshooting relay out-
puts.
-
Verify 5X00 wiring and configuration.
Troubleshooting
Sensor Troubleshooting continued
Issue
possible cause
possible solution
relay does not energize
incorrect sensor configura-
tion
See
Sensor Set up
relay not enabled
incorrect timer configuration See
Timers
30 second delay at power up
and/or sensor configuration
change
30 seconds delay - see
248..
peripheral device is wired
incorrectly
See
Wire Relays
device is incorrectly wired
to NO or NC terminal strip
ports..
relays are not enabled for
control/alarm/timer system
Verify menu configuration
(relay enable) for control/
alarm/timer system.
See sensor system set up -
150.
See timer system enable
-180.
temp (high) does not de-
energize when set point
value is read
sensor is configured for
timed control
Verify setpoint and control
configuration values -
Email Alarm Troubleshooting
Issue
possible cause
possible solution
no 5X00 email alarm
message received
5X00 display is not at Run
Screen - all alarming is sus-
pended when in menus
exit to run screen to reset
alarm email alarm system
Serial to ethernet device did
not initialize properly. Check
the event log for “unable to
initialize error”.
disconnect power to 5X00,
verify installation of serial
to ethernet device on lower
board and repower
faulty ethernet serial to eth-
ernet device
replace serial to ethernet
device
Troubleshooting