YSI 5X00
171
YSI 5X00
170
2. Wire relay output device(s) to local instrument, or any networked 5X00
instrument - page 54.
OR
2. Wire relay output device to REM - see IOEM/REM Safety and Instal-
lation guide>Chapter 3>Step 4>Wire REM Relay Outputs.
3. Configure “temperature” input type (for non ODO sensor temperature
system) at
Menus → System → Configure Aux →Input Type
4. Configure temperature source :
Menus → Sensor Setup→ DO 1-4→Temp Source→User Defined
→Aux
3-6
OR
M
enus → System → Configure (Aux 3-6 only) →Temp Source →Aux Input
→ODO Probe
5. Enter Aux 3-6 in Sensor Setup menu to configure temperature system:
Menus → System → Sensor Setup→Aux 3-6 (Temperature 1-6)
6. Configure Control Mode see 152.
Menus → System → Sensor Setup
→Set Point Control
→PID/PWM Control
7. Use
▲
▼
to scroll and highlight submenus.
5400
5500
Configuring the 5X00
8.
Press
to select.
9. Enable and configure submenus. See Control Modes and Alarms for
menu functionality - page 152. Use AquaManager to configure sensor
setup menus - page 150.
10. Optional - Configure digital 4/20mA output - page 178.
11. Press
to save settings and return to Run Screen.
12. Test system output devices at→System→Service→Output Test - page 144
or create an event to verify configured relay(s) energize.
Configure Temperature System E-Aux 1-8
Before configuring E-Aux temperature systems 1) See - “Use AquaManager to
configure system inputs and outputs” - page 150 AND 2) See Control modes
[Set point Control Mode and Alarms - page 153 (E-Aux temperature set point
control mode can operate in normal or timed mode - page 157) and PID control
mode - page 159.]
1. Wire temperature input at IOEM for E-Aux system - see IOEM/REM
Safety and Installation guide>Chapter 3>Step 5>Wire IOEM inputs;
2. Wire relay output device to local instrument, or any networked 5X00
instrument - page 54 for local instrument.
OR
2. Wire relay output device to REM - see IOEM/REM Safety and Instal-
lation guide>Chapter 3>Step 4>Wire REM Relay Outputs. Wire any
output devices to be used for system.
3. Configure Input Type at
Menus → System → Configure E-Aux→Input
Type and other fields
- page 108.
4. Enter E-Aux 1-8 in Sensor Setup menu
Menus → System → Sensor Setup → E-Aux 1-8
5. Use
▲
▼
to scroll and highlight submenus.
6.
Press
to select.
7. Enable and configure submenus. Configure IOEM Source and Chan-
nel. Also see Control Modes and Alarms for menu functionality - page
152. Use AquaManager to configure sensor setup menus - page 150.
8. Optional - Configure digital 4/20mA output - page 178.
9. Press
to save configuration and return to Run Screen.
10. Test system output devices at→System→Service→Output Test - page 144
or create an event to verify configured relay(s) energize.
Notes:
-
It is important to keep the 5X00 at the Run Screen during normal opera-
tion. Alarm functionality (email alarms, activation of enabled alarm relays,
alarm icons, and buzzer) is suspended when 5X00 is not at Run Screen.
E-AUX
E-AUX
Configuring the 5X00