Configure DO Systems
Before configuring E-Aux temperature systems 1) See - “Use AquaManager to
configure system inputs and outputs” - page 150 AND 2) See Control modes
[Set point Control Mode and Alarms - page 153 (E-Aux temperature set point
control mode can operate in normal or timed mode - page 157) and PID control
mode - page 159.]
1. Install and wire DO probe(s) and membrane and/or sensor cap - 49.
2. Wire relay output device to local instrument, or any networked 5X00
instrument - page 54.
OR
2. Wire relay output device(s) to REM - see IOEM/REM Safety and Instal-
lation guide>Chapter 3>Step 4>Wire REM Relay Outputs.
3. Configure DO system temperature source :
Menus → Sensor Setup→ DO 1-4→Temp Source→User Defined
→Aux
1-6
5400
OR
Menus → System → Configure (Aux 3-6 only) →Temp Source →Aux Input
→ODO Probe
4. Configure Control Mode - 152.
Menus → System → Sensor Setup→Set Point Control
→PID/PWM Control
5. Use
▲
▼
to scroll and highlight submenus.
6.
Press
to select.
7. Enable and configure submenus. See Control Modes and Alarms for
menu functionality - page 152. Use AquaManager to configure sensor
setup menus - page 150.
8. Optional - Configure digital 4/20mA output - page 178.
9. Press
to save settings and return to Run Screen.
10. Test system output device(s) at→System→Service→Output Test - page
144 or create an event to verify configured relay(s) energize.
Note:
-
Elevation data can only be viewed and configured locally at the 5500D instru
ment and cannot be viewed or configured using AquaManager.
PID 4/20mA and PID/PWM control mode menus at instrument
All non digital sensor systems can be configured for PID/PWM or PID 4/20mA
control mode.
See PID/PWM and PID 4/20mA Control Mode - page 175 before configuring
sensor systems.
5500
5500
Configuring the 5X00
Configuring the 5X00