39
XXX To R XXX ".
If the electrode's position is to the right, the screen displays "Gap
XXX To L XXX".
If the electrode's deviation is within 10, no adjustment is needed. The
electrodes location adjustment requires high skill. It should be done
by a technician. If the value of "GAP" is between 15 and 25, the arc
power is suitable for fiber splicing. If the value of “GAP” is less than
this range, please increase "Prefuse Power" and "Fuse Power",
contrarily, decrease "Prefuse Power" and "Fuse Power". Press
the
key to return to the former operation after the test is done.
Before the Arc Power test, the electrodes should be cleaned. The test
is effective only after the arc becomes stable.
A.3 Overlap Test
Choose the "Overlap Test" item in test menu and press
key to
confirm. Now the screen displays as Figure A-3.