8.1.3 Cleaning the Objective Lens
The surface of the objective lens must be kept clean. Dust on
lens can cause trouble in the image process and a big splice
loss.
(1) Turn off the power.
(2) Use a cotton swap soaked with pure alcohol to clean the
lens’s surface from center toward rim.
(3) Turn on power. Make sure dust cannot be seen on the LCD
screen.
8.2 Cautions
(1) The splicer is designed for optical telecommunication
fibers. Don't insert any other type of object in the fiber
clamps or arc center.
(2) Make sure the main body of the splicer is grounded during
use.
(3) During discharge, there is a high voltage of several
thousand voltages between the electrodes. So don’t touch
them. If there are drops of water on the splice’s main body,
dry it thoroughly before use.
(4) When alcohol drops (including hand grease etc) are
adhered to the electrodes, accidental discharge might occur.
However, after discharging several times, such liquids will
disappear and normal discharge will proceed.
(5) Do not use the splicer in the environment full of