aquaMAX
TM
– Ultra 370 Series
- 25 -
[
Fig. 4-3
]
UF Filter
○
4
Detach 2 tubes from inlet and outlet of UF filter. There could be produced water left in tubes.
○
5
After detaching all tubes connected to the filter, take off the filter from the holder.
○
6
Install a new one, then connect tubes to inlet and outlet.
○
7
Press
Service
key and select “Stabilization”. Press, then,
Product/Recirculation
key to run it.
- Get a reservoir tank for produced water. Get another tank for REJECT water.
- Perform it 2~3 times to remove air completely.
○
8
Check for the leakage during production. Make sure the resistivity of produced water gets 18.2M
Ω·
cm