aquaMAX
TM
– Ultra 370 Series
- 24 -
(c) Detachment of a cap from UV chamber (d) Taking out of UV lamp
[
Fig. 4-2
]
UV lamp chamber
○
12
Press
Service
key and select “Stabilization”. Press, then,
Product/Recirculation
key to run it.
- Get a reservoir tank for produced water.
- Perform it 2~3 times to remove air completely.
○
13
Check for the leakage during “Stabilization”.
○
14
After checking all functions work well, start to produce water.
- Make sure the resistivity of produced water gets 18.2M
Ω·
cm
3) UF Filter Replacement.
○
1
Press
Service
key on the standby mode.
○
2
Press
Service
key 6 times, and Product key 2 times to perform CHANGE FILTER.
○
3
After it displays OK, it turns to Standby mode, then, open the front door.