aquaMAX

TM

 – Ultra 370 Series 

 
 

 

- 14 - 

Chapter 3. Operation 

3-1.Key Operations 

Display

 

 - The resistivity of produced water is shown. Press 

Display

 key during operation to see the temperature. 

Press once more to get back the resistivity display. 

Service

 

- This key has various functions up to 6 (or 7) on 

Standby Mode

. Press 

Service 

key to move to each 

function and press 

Production/Recirculation

 key to run. 

 

1. Stabilization 

It is to remove air in filters and tubes at the first installation or replacement. The system produces the 

produced water for 1 min. ( The REJECT valve is opened for 10 sec, and waste water will be drained through 

the REJECT line at the back-side. 

 

 

S E R V I C E

 

 

S t a b i l i z a t i o n . .

   

 

 

 

S E R V I C E

 

 

C o m p l e t e d. .

 

 

 

 

2. Press Down 

It’s used to reduce inlet pressure when any parts of filters need to be replaced.( Waste water will be drained 

through the REJECT line at the back-side. 

 

 

S E R V I C E

 

 

P R E S S    D O W N

 

 

 

Summary of Contents for aquaMAX Ultra 370 Series

Page 1: ...of this manual are subject to change without notice in accordance with product improvements This operation manual describes the operation over the life of this instrument carefully read this manual t...

Page 2: ...Installation 9 2 1 Packing 9 2 2 Installation Guide 10 2 3 Installation Requirements 12 2 2 1 Installation Requirements 12 2 2 2 Safety Indication 12 2 2 3 WEED GUIDE 12 3 Operation 14 3 1 Key Operati...

Page 3: ...M Ultra 370 Series Accessories 6 Fig 2 1 The Configuration of the aquaMAXTM ultra 9 Fig 2 2 Installation of Ultra Pack 10 Fig 2 3 Installation of Final Filter 10 Fig 2 4 Tube connections for Feed Wate...

Page 4: ...V Ultraviolet lamp for low TOC level 5ppb and a UF Ultrafilter for Pyrogen free lower than 0 06 Eu ml This system uses feedwater from R O purification system and the produced ultrapure water complies...

Page 5: ...is to remove ion from an ion exchange resin Ion exchange capable materials have week chemical combination The materials have metathesis property a positive ion by a positive ion negative ion by a nega...

Page 6: ...5 ppb 1 2 3 UF Filter UF filter which is made from permeate hard and thin membrane reduces almost of large molecular It is also effective to reduce most particles and bacteria Pyrogen level to be 0 06...

Page 7: ...aquaMAXTM Ultra 370 Series 4 1 3 Flow Diagram UV lamp and UF filter are optional accessories Fig 1 2 Flow Diagram...

Page 8: ...y Display of temperature during the production Service Selectable among the Flush Press down Filter Change Operate Standby Change the operate standby state Production Recirculation Change the Producti...

Page 9: ...Ultra 370 Series 6 Rear side Fig 1 4 aquaMAXTM Ultra rear side Accessories ULTRA Pack Final Filter UF Filter UV Lamp power switch power connector RS232C connector Fig 1 5 aquaMAXTM Ultra 370 Series Ac...

Page 10: ...1 Heavy Metal ppm 1 0 05 0 01 0 01 0 1 0 01 Dissolved Organics ppm 1 1 1 1 Bacteria cfu 100 10 100 10 Small Particles 5 10 000 100 10 000 10 Table 1 2 Water Classification According to CAP ASTM Standa...

Page 11: ...LTRA Pack High purity cartridge 0 2 m Final filter Flow Rate 0 5 1 5 L min Depending on Water Pressure Resistivity 18 2 M cm 25 TOC Pyrogen Particle Micro organic Silicate Heavy metals Water Analysis...

Page 12: ...ollowing items Fig 2 1 The Configuration of the aquaMAXTM ultra 1 aquaMAXTM Ultra system fuse 250V 1A mounted on the fuse holder 2 ULTRA Pack cartridge 3 0 2 m Final Filter 4 Tubings for the feed wate...

Page 13: ...d Press down the clip connected to the fitting making a sound like click Insert the filter to fix into the marked fitting by making click sound There is an indicating mark at the lower part of filter...

Page 14: ...Feed Water Connection Reject Water Connection Connection Completed Fig 2 4 Tube connections for Feed Water and Reject Water Use for R O Reverse Osmosis grade water or D I Deionized water as feed wate...

Page 15: ...equipment may be impaired 2 2 2 Safety Indication An explanation of symbols related to safety which are used on the equipment a Protective Earth b and Power On and Power Off c Caution risk of electric...

Page 16: ...roduct or their local government office for details of where and how they can take this item for environmentally safe recycling Business users should contact their supplier and check the terms and con...

Page 17: ...ction Recirculation key to run 1 Stabilization It is to remove air in filters and tubes at the first installation or replacement The system produces the produced water for 1 min The REJECT valve is op...

Page 18: ...Press Service key 3 times To view the conductivity run as follows S E R V I C E DISP RESIST S E R V I C E DISP CONDUCT 4 Sanitize It is to regularly wash or install a UF filter S E R V I C E S a n i t...

Page 19: ...below S E R V I C E R E A D Y T O C H A N G E Press Production Recirculation key it reduces inlet pressure and turns back to the Ready mode Then change filters S E R V I C E S E R V I C E P R E S S D...

Page 20: ...ulation is stopped LCD still displays Operate mode if the resistivity is higher than 13M cm production is available any time by pressing PRODUCTION key Pressing OPERATE STANDBY key during operation ge...

Page 21: ...elow STANDBY Ready On the Ready mode the system runs Recirculation to prevent filter from contamination every 1 hour for 5 minutes STANDBY 5min Recirculation System configuration aquaMAX ultra 370 Bas...

Page 22: ...cm after FLUSH run it again OPERATE S Y S T E M C H E C K 3 Press the Product Recirculation key to produce If the purified water quality does not reach to 13 M cm following message is displayed and th...

Page 23: ...backs to Standby mode Normal operation is still available but To replace filters run CHANGE FILTER function on Service menu CHANGE FILTER message keeps popping up every 30 min O p e r a t e STANDBY U...

Page 24: ...the system keeps power on it can increase the time of use because recirculation runs every 1 hour 4 2 Filter safekeeping For Ultra Pack or UF Filter If the system is not used for a long time keep the...

Page 25: ...all amount of water leaking during the filter replacement Remove the cap The cap for input and output of filters then install Ultra Pack 1 and 2 each by no on the instrument Insert it to fix into the...

Page 26: ...lamp power code on the upside of UV chamber See the Fig 4 2 c Carefully lift up the cap in straight with UV chamber as following Fig 4 2 d Grab the connector cap with one hand and carefully detach the...

Page 27: ...ced water Perform it 2 3 times to remove air completely 13 Check for the leakage during Stabilization 14 After checking all functions work well start to produce water Make sure the resistivity of prod...

Page 28: ...off the filter from the holder 6 Install a new one then connect tubes to inlet and outlet 7 Press Service key and select Stabilization Press then Product Recirculation key to run it Get a reservoir ta...

Page 29: ...ata such as temperature and resistivity through RS232C connector In the case of request to Younglin Instrument receive data on txt cable Part No ___ connect to external computer by Terminal VT100 Data...

Page 30: ...How to calibrate On the Recirculation Production mode display Resistivity Temperature Data acquisition by connecting with external computer is available Measure the resistivity measurement by resisti...

Page 31: ...n the final filter 2 Run the stabilization and then water get drained through the final filter on production or Purge air mode 3 Open the valve of final filter to remove air inside on Production mode...

Page 32: ...ower on the unit 7 Push the SERVICE key four times then Push the Production key to select the Sanitize Unit 8 Inject solution state will be displayed then open the UF sanitization port 9 Fill syringe...

Page 33: ...te key to enter to try again the first sanitize 14 The sanitizing procedure must be continued until the state has stable value 15 If the state shows stable value 1 the entire sanitization procedure is...

Page 34: ...push the Operate key After the Second Cleaning Sanitizing is operated for 20 minutes and then Feed water is displayed If PE reservoir is filled feed water more than 20L push the Operate key After the...

Page 35: ...3 3701011003 Accessories ULTRA Pack 3501012070 0 2 m Final Filter 1604023641 UV Sterilizing filter 1604023660 UV lamp 1604023620 UF filter 1604023651 Tubings for feedwater produced water 35030100 Powe...

Reviews: