Quick Setting Function
5
Setting Example
Set the following parameters to set to high limit control and thermocouple Type K (range:
0.0 to 500.0ºF). No need to change the parameters other than the following parameters.
Set QSM = YES to enter the quick setting mode.
(1) Set HI.LO = HIGH.
(2) Set IN = K1.
(3) Set UNIT = F.
(4) Set RH = 500.0.
(5) Set RL = 0.0.
Set EXIT = YES to quit the quick setting mode.
The Operation Display is shown.
Setting Details
Control Type
Parameter
symbol
Name
Display
level
Setting range
Menu symbol
HI.LO
Limit control type
EASY
LOW: Low limit control
HIGH: High limit control
CTL
5.1 Setting Using Quick Setting Function