<APPENDIX 1. LIST OF PARAMETERS FOR EACH BLOCK OF RAMC>
A-6
All Rights Reserved. Copyright © 2006, Rota Yokogawa
IM 01R01B02-01E-E 5th edition October 01, 2020-00
Abs.
Index
Parameter
Sub-
Parameter
Label
Data Type/
Structure
Read/
Write
Initial value
Functional Description
48
LO_ALM
-
DS-39
R/-
-
State of the lower warning limit
Unacknowledged
Lower Unack.
Warnings
Unsigned8
R/-
0
For future use (not available)
Alarm_State
Lower Warning
Status
Unsigned8
R/-
0: No Warning
Indicates the warning status
Time Stamp
-
Time Value
R/-
0
For future use (not available)
Subcode
-
Unsigned16
R/-
0
For future use (not available)
Value
Lower Warning
Output Value
Float
R/-
0
Indicates the value which caused the
warning
49
LO_LO_ALM
-
DS-39
R/-
-
State of the lower alarm limit
Unacknowledged
Lower Unack.
Alarms
Unsigned8
R/-
0
For future use (not available)
Alarm_State
Lower Alarm
Status
Unsigned8
R/-
0: No Alarm
Indicates the alarm status
Time Stamp
-
Time Value
R/-
0
For future use (not available)
Subcode
-
Unsigned16
R/-
0
For future use (not available)
Value
Lower Alarm
Output Value
Float
R/-
0
Indicates the value which caused the
alarm
50
SIMULATE
-
DS-50
R/-
-
Simulation of Transducer Block input
into the Analog Input Function Block
Simulate_Status
Simulation Status
Unsigned8
R/-
-
Indication and input of simulation value
Simulate_Value
Simulation Value
Float
R/-
0.0
Indication and input of simulation status
Simulate_Enabled Simulation
Unsigned8
R/-
0: Disabled
Enables and disables simulation (inter-
rupts transducer block input)
51
OUT_UNIT_TEXT
Out unit text
Octet String
R/-
-
Indication and input of customer specific
textual unit of OUT parameter (used for
units not contained in the code list).
230
VIEW_1
-
R/-
n.a.
Indicates a Profile defined parameter
preselection
231
VIEW_2
-
R/-
n.a.
Indicates a Profile defined parameter
preselection