<2. Specifications>
2-4
IM 12F4A1-01E
Air purge (using instrument air)
Supply pressure: 140 kPa
Air consumption: Approx. 5 l/min
Weight
For pure water:
Approx. 65 kg
For raw water (1-cylinder): Approx. 70 kg
For raw water (2-cylinder): Approx. 75 kg
Optional specifications
Pump for air purge (if instrument air cannot be used)
Install to enable intake of clean air.
Intake/discharge:
5 l/min, at 50 Hz (Intake/discharge outlets: Open)
Maximum pressure: 80 kPa
Power consumption: Approx. 23 VA
Weight:
Approx. 2 kg
Regulatory Compliance:
EMC Regulatory Arrangement in Australia and New Zealand (RCM)
EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard Class A
한국 전자파적합성 기준
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
n
Characteristics
(Percent display is computed with respect to whichever of output range 1 or output range 2 has
the highest upper range value)
Repeatability:
2%
Linearity:
±3%
Drift:
Zero drift:
±1%/month Max.
Span drift:
-5%/month Max.
Response time:
Displayed as 90% response time
For pure water:
Approx. 3 min
For raw water:
Approx. 4 min (Within 3 min from standard liquid inlet)
Temperature compensation error (water temperature):
±1% Max. (Temperature compensation range: 0 to 40 °C)
Effect of ambient temperature: ±0.5%/10 °C
Effect of power variation:
±0.5%/10% of rated voltage
Effect of combined chlorine: 5% or less, of concentration of combined chlorine for combined
chlorine insensitive version
14th Edition : Jul. 15, 2020