![YOKOGAWA RC400G User Manual Download Page 115](http://html.mh-extra.com/html/yokogawa/rc400g/rc400g_user-manual_911334115.webp)
<8. Maintenance>
8-14
IM 12F4A1-01E
8.10.5
Replacement of Activated Charcoal Filter
Stop supplying sample water by closing the main valve.
1. Remove the filter case by turning it in the direction shown below.
2. Clean the inside of the case with a brush or appropriate tool.
3. Replace the filter with a new one.
4. Reverse the above steps to assemble.
5. Turn on the solenoid valve SV6 to allow chlorine-free water to run for at least 30 minutes.
Check that there is no leakage.
6. Check that the reading approximates to zero.
Turn
Cover (fixed by pipe)
Filter (activated charcoal)
Spacer
Filter Case
Figure 8.9
Replacing Activated Charcoal Filter
14th Edition : Jul. 15, 2020
Summary of Contents for RC400G
Page 5: ...Blank Page ...
Page 9: ...Blank Page ...
Page 25: ...Blank Page ...
Page 41: ...Blank Page ...
Page 95: ...Blank Page ...
Page 139: ......