<5. Converter Operation>
5-16
IM 12F4A1-01E
Diffusion current
(µA)
6
5
4
3
2
1
0
0
-0.5
-1.0
Step 0
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Plateau characteristic (example)
Applied voltage (V)
Note
The applied voltage changes in the direction of increasing negative voltage.
The number of diffusion current samples is “the number of steps plus one”.
Applied voltage at each step are computed automatically when executed.
T6.4.ai
(Setting example)
Starting voltage
: 0 V
Stopping voltage : -0.8 V
Number of steps
: 8
Relevant to initial values
Starting
voltage
Stopping
voltage
Figure 5.4
Example of Taking Plateau Characteristics
The key operation procedure for <taking plateau characteristic> is as follows:
(Example)
(a) Select function code “9" in <MAINT.> mode.
→ PLt_E
(Display)
(b) On pressing [ENT] key, the starting voltage is displayed. → L00.00 V
(c) To change the setting, press [>] key.
→ L
0
0.00 V
(If no setting change is required, operate from step (e).)
(Leftmost digit flashes.)
(d) Select a digit using [>] key and set the numeric value
using [Λ] and [>] keys.
→ L00.
1
0 V
Pressing [ENT] key now fixes the numeric value.
(Last set digit flashes.)
(e) Press [ENT] key to display the stopping voltage.
→ H-1.00 V
(f) To change the setting, press [>] key.
→ H
-
1.00 V
(If no setpoint change is required, start from step (h).)
(Thenegative sign flashes.)
(g) Select a digit using [>] key and set the numeric value
using [Λ] and [>] keys.
→ H-0.
8
0 V
By next pressing [ENT] key, the numeric value is fixed.
(Last set digit flashes.)
(h) On pressing [ENT] key, the number of steps is displayed. → P0010
(i) To change the setting, press [>] key.
→ P
0010
(If no setpoint change is required, start from step (k).)
(Leftmost numeral flashes.)
(j) Select a digit using [>] key and set the numeric value
using [Λ] and [>] keys.
→ P000
8
Pressing [ENT] key now fixes the numeric value.
(Last set digit flashes.)
(k) Pressing [ENT] key displays "Execution selection".
→ Go_oF
If each setpoint is to be reconfirmed, press [ENT] key.
The display returns to (b) starting voltage display.
(l) If the displayed item is to be executed, press [>] key.
→ Go_on
(m) On pressing [ENT] key, characteristic taking is executed.