<2. Specifications>
2-5
IM 12F4A1-01E
n
Product model and suffix codes
[Style: S4]
Model
Suffix code
Option
code
Description
RC400G
--------------------------------- ---------- Residual chlorine analyzer
Application
(cleaning unit)
(*1)
-1
---------- For pure water, for municipal water (with glass-bead
cleaning)
-2
---------- For raw water (1-cylinder) (with glass-beads cleaning, jet
cleaning, and one cylinder sand filter unit) (*6)
-3
---------- For raw water (2-cylinder) (with glass-beads cleaning, jet
cleaning, and two cylinder sand filter unit) (*6)
Output signal
5
----------
1 to 5 V DC
6
---------- 4 to 20 mA DC
Power supply
1
----------
200 V AC, 50 H
z
2
----------
200 V AC, 60 Hz
3
----------
220 V AC, 50 Hz
4
----------
220 V AC, 60 Hz
5
----------
100 V AC, 50 H
z
6
----------
100 V AC, 60 H
z
7
----------
110 V AC, 50 Hz
8
----------
110 V AC, 60 Hz
Measurement object
(*2)
F
---------- Free chlorine
T
---------- Residual chlorine (total chlorine)
C
---------- Combined chlorine insensitive version (free chlorine) (*3)
-
-N
---------- Always -N
-
*A ---------- Always A
Optional specifications
/AP1
With air purge pump (200 V AC, 50/60 Hz) (*7)
/AP3
With air purge pump (220 V AC, 50/60 Hz) (*7)
/AP5
With air purge pump (100 V AC, 50/60 Hz) (*7)
/AP7
With air purge pump (110 V AC, 50/60 Hz) (*7)
/PPM
Units: ppm
/NR
Without reagent set for start-up (*4)
/SCT
Stainless tag plate
/S
Stainless stand (*5)
/AZC
With automatic zero calibration
/ARS
With arrestor (power and signal lines)
*1:
“Application” indicates general guidelines. Select the cleaning equipment appropriate for the contamination in the sample.
*2:
Measurement object selection is indicated in the following.
*3:
Less affected by combined chlorine in free chlorine measurement.
*4:
When ordering the RC400G, select /NR and get reagent sets for start-up from local reagent supplier.
These reagent sets can not be exported from Japan due to both safety and transportation issues.
*5:
Stanchion, base and bracket are stainless steel.
*6:
Please contact Yokogawa regarding the adequate number of the cylinder.
*7:
Installation of air purge is always required for all application. If instrument air cannot be used, select the air purge pump.
14th Edition : Jul. 15, 2020
Summary of Contents for RC400G
Page 5: ...Blank Page ...
Page 9: ...Blank Page ...
Page 25: ...Blank Page ...
Page 41: ...Blank Page ...
Page 95: ...Blank Page ...
Page 139: ......