27
IM 77C01E31-01EN
SFLC-212-181
27
5.2
Setting table
A setting item changes by the specification of a product.
(1) Important setting
Each parenthesized number shows a setting number and this number is displayed on the setting screen.
Items
Setting and operation procedures
Page
Sets the phase wire
of input circuit.
(311)
Press SET and DISPLAY together for lo
nger than 3 seconds
(311)
Select an phase wire by
+
and
-
Press SET
for longer than 3 seco
nds
Selected phase wire is entered
Press DISPLAY
Returns to display mode.
50
Sets the input
voltage
(phase-voltage full
scale).
(312)
Press SET and DISPLAY together for
longer than 3 seconds
Press
RESET/SHIFT
(311)
(312)
Select an input voltage (In case of 1
φ3W, it is phase
-voltage full scale) by
+
and
-
Press SET
Selected input voltage is entered
Press DISPLAY
Returns to display mode.
51
Sets the
measurement range
of voltmeter
(211)
Press SET and
RESET/SHIFT
together for longer than 3 seconds
(211)
Select a measurement range by
+
and
-
Press SET
Selected measurement range is entered
Press DISPLAY
Returns to display mode.
43
to 46
Sets the
measurement range
of ammeter
(213)
Press SET and
RESET/S
HIFT
together for longer than 3 seconds
(211)
Press
RESET/SHIFT
Press
RESET/SHIFT
Select a measuring range by
+
and
-
(212)
(213)
Press SET
Selected measuring range is entered
Press DISPLAY
Returns to display mode.
43
to 46
Sets the display
combination
(111)
Press SET
for longer than 3 seconds
Select the display combination by
+
and
-
(111)
Press SET
Selected display combination is entered
Press DISPLAY
Returns to display mode.
35
to
39
Set the address of
device in
communication
output.
(221C)
Press SET and
RESET/SHIFT
together for longer
than 3
seconds
Press MODE
(211)
(221C)
Select an address by
+
and
-
Press SET
Selected address is entered
Press DISPLAY
Returns to display mode.
47
Set the
transmission rate of
communication
output.
(222
C)
Press SET and
RESET/SHIFT
together for longer than 3 seconds
Press MODE
(211)
(221C)
Press
RESET/SHIFT
Select an transmission rate by
+
and
-
Press SET
(222C)
Selected transmission rate is entered
Press DISPLAY
Returns to display mode.
47
Set the parity bit
added to
communication
data.
(223C)
Press SET and
RESET/SHIFT
together for longer than 3 seconds
Press MODE
(211)
(221C)
Press RES
ET/SHIFT
Press
RESET/SHIFT
Select an parity by
+
and
-
(222C)
(223C)
Press SET
Selected parity is entered
Press DISPLAY
Returns to display mode.
47