8-20
<Toc> <Ind>
<8. PR201 Original Communication Protocol>
IM 77C01E01-10E
●
Parameter: Z (Error response)
• Function
Returns an error response.
• Command/response
Number
of bytes
1
2
1
2
2
1
1
Command
element
STX
DG
Z
Station
number
(ADR)
Checksum ETX
CR
There is no command data for data reading.
1
2
1
2
2
Response
element
STX
DG
Z
Station
number
(ADR)
ⵧⵧ
2
1
1
Checksum ETX
CR
Number
of bytes
Response data is returned in a 2-byte ASCII character string.
• Example
Reads an error response of PR300 with station number 01.
䊐䊐
indicates the checksum
value.
[Command]
[STX]DGZ01
䊐䊐
[ETX][CR]
Returns a checksum error response for the above command.
[Response]
[STX]DGZ0180
䊐䊐
[ETX][CR]
Reads an error response by converting hexadecimal data “80” into binary data
“10000000” and then reading the 7th bit (checksum error).
• Bit information for error response
Bit
Failure Status
0
Overranged power
1
Undefined
2
Undefined
3
Undefined
4
Undefined
5
Command data format error
6
Overranged command data
7
Checksum error
Summary of Contents for PR300