1. Overview
37
TI 11B08A01-01E
l
Piping connection of control unit, isothermal oven, large isothermal oven, and
programmed temperature oven
15
30
23
139.5
230
30
15
30
23
155
230
30
39
20
170
23
138
138
23
170
20
39
20.5
40
Unit :mm
Isothermal
oven
Large
isothermal oven
Air outlet for stream switching valve
6mm or 1/4” tube
Protective gas (instrumental air) inlet
Rc1/4 or 1/4NPT
Protective gas (instrumental air) inlet
Rc1/2 or 1/2NPT
Vent for protective gas (instrumental air)
Inlet/outlet for sample gas, carrier gases, etc.
6mm or 1/4” tube
Control unit
Programmed
temperature
oven
l
Analyzer base sampling system (GCSMP)
120
215
55
120
60
30
60
150
11
0
165
80
120
30
300
60
30
60
150
11
0
165
80
215
55
60
*: It depends on specifications.
For Type 1
Steam inlet *
Rc1/4 or 1/4NPT
Steam outlet
or drain *
Rc1/4 or 1/4NPT
Inlet/outlet for sample
gas, standard gas, etc.
Rc1/4 or 1/4NPT
For Type 2, Type 4
Steam inlet *
Rc1/4 or 1/4NPT
Steam outlet
or drain *
Rc1/4 or 1/4NPT
Inlet/outlet for sample
gas, standard gas, etc.
Rc1/4 or 1/4NPT
Unit: mm
*
Some specifications do not have these connections.
Sep. 07, 2017-00