9
TI 11B08A01-01E
n
General Precautions
WARNING
In order to analyze gases, process gas chromatographs use various kinds of process gases and
utility gases.
Since these gases are typically combustible, combustion-sustaining, toxic, odorous, resolvable,
polymerizing, or corrosive, refer to the “Safety Information” in our approval drawings and others
to ensure safety thoroughly before using them.
WARNING
• The GC8000 weighs about 100 to 220 kg. Unpack it near the installation site. Use a
transportation machine to move it. Handle it carefully to prevent it from falling.
•
Up to two protection system may be included, each of which weighs approximately 7 kg, are
installed on top of the GC8000. Therefore, the center of gravity is higher than the center of
the analyzer body.
CAUTION
For lifting and carrying the equipment, use those two (or three) fixture brackets which are fixed to
the wooden (or other materials) crate for the equipment.
CAUTION
•
Since the GC8000 is a precision instrument, take care when handling it to avoid impact.
• The GC8000 is designed for use under specified environment (over voltage category,
pollution degree, ambient temperature, flowrate and pressure of sample and utility gas and
specification of electric power). It is out of warranty against occured result in case when
exceed these specs.
• Use the GC8000 within the range of your purchase specifications.
Yokogawa assumes no responsibility for problems resulting from use by the customer
outside the purchase specifications.
If the GC8000 needs to be modified or repaired, please contact your nearest Yokogawa
representative. Yokogawa assumes no responsibility for results where the customer or any
third party has attempted to modify or repair this product.
•
For cleaning of analyzer, do not use any abrasives or organic solvent.
CAUTION
This instrument is a Class A product for EMC standard (Emission), and is designed for use in an
industrial environment. Please use this instrument in an industrial environment only.
Complying with EMC Standard for ATEX, IECEx, and NEPSI:
•
EN61326-1 Class A (Emission)
•
EN61326-1, EN61326-2-3 (Immunity)
• Korea Electromagnetic Conformity Standard
Sep. 07, 2017-00