1. Overview
42
TI 11B08A01-01E
(6) FID/FPD combustion air
This air is used to burn hydrogen gas in an FID or FPD.
The air must satisfy the following conditions.
Purity:
Measuring range from 0 to 50 ppm or more: moisture: 10 ppm or less; organic
components: 5 ppm or less
Measuring range from 0 to less than 50 ppm: moisture: 5 ppm or less; organic
components: 0.1 ppm or less
(7) Steam
Steam is required to steam-heat a sample. Prepare a steam source that can apply the pressure
listed in “Operation Data.”
1.5
Conformance Standards
Safety Standard, EMC Standard and RoHS Standard are showned in the following list.
Ex Certification
MS code
Safety Standard
EMC standard
RoHS standard
TIIS
GC8000-T
-
-
-
ATEX *1
GC8000-A
EN 61010-1
EN 61010-2-030
GB 30439.1 (for NEPSI)
EN 61326-1 Class A, Table 2
EN 61326-2-3
Korea Electromagnetic
Conformity Standard
RCM Mark
EN 50581
IECEx *2
GC8000-E
-
NEPSI *3
GC8000-P
FM
GC8000-F
GC8000-G
FM 3810: 2005
(ANSI/ISA 61010-1-2004
(82.02.01),
ANSI/ISA-82.02.02-1996
(IEC 61010-2-031))
-
-
CSA
GC8000-C
GC8000-D
CSA C22.2 No. 61010-1-04
-
-
*1: Analyzer base sampling unit (GCSMP) or base placed under GC8000-A is not covered by the EU-DoC nor
the EC Type-examination Certificate. The empty compartment (base) placed by Yokogawa to create a Self-
standing GC8000-A, does not impair the compliance of the GC8000-A.
*2: The design is based on the safety and EMC standard, though the mark of CE is not indicated.
*3: The design is based on the safety and EMC standard, though the mark of CE, KC and RCM are not
indicated.
CAUTION
This instrument is a Class A product, and is designed for use in an industrial environment. Please
use this instrument in an industrial environment only.
A list below shows the applicable standards for electrical equipment for explosive atmospheres.
This list does not include standards other than related to electrical equipment for explosive
atmospheres.
For more details, see each certificate. Please contact your nearest Yokogawa representative.
Ex Certification
MS code
Standards for electrical equipment for explosive atmospheres
TIIS
GC8000-T
Constructional requirements for electrical equipment for explosive
atmospheres
Reference guide; Type examination guide for electrical equipment for explosive
atmospheres (relating to technical standards conforming to international
standards) issued by the TIIS (Technical Institution of Industrial Safety), Japan.
(Issued in November, 1996.)
ATEX
GC8000-A
EN 60079-0:20012+A11, EN 60079-1:2014, EN 60079-2:2014
IECEx
GC8000-E
IEC 60079-0:2011, IEC 60079-1:2014, IEC 60079-2:2014
NEPSI
GC8000-P
GB 3936.1-2010, GB 3836.2-2010, GB 3836.5-2004
Sep. 07, 2017-00