2. Installation, Piping, and Wiring
75
TI 11B08A01-01E
CAUTION
The minimum packing (for ø9.0 to ø10.0) is attached to the cable packing adapters at shipment.
Change it to an appropriate packing for the cable O.D. (See Table 2.7.)
Hex.socket set screw
Cable packing adapter
Nut
Connection threads
Case of electronics section
(Pressurized enclosure)
Gasket
Body
Identification mark
Double-deck packing
(Replace it with an appropriate one from the accessory kit.)
Guide ring
Gland
Coupling
Cover
Cable
Hex key wrench
Hex.socket set screw
Plate
This plate is included with shipment.
Please keep the removed the plate.
To the connection port of
the electronics section
Figure 2.18
Procedure for mounting a cable packing adapter for the electronics section
The cable packing adapters comply with the Technical Standard of the Ministry of Health, Labour
and Welfare, Japan.
Table 2.7
Double-deck packing size for electronics section
Connection
port screw
Applicable
cable O.D.
Identification mark
Recommended
torque for the
gland (N·m)
G3/4
ø9.0 to ø10.0 SCFP2010 ø9.0 to ø10.0
25
ø10.0 to ø11.0 SCFP2011 ø10.0 to ø11.0
20
ø11.0 to ø12.0 SCFP2012 ø11.0 to ø12.0
20
ø12.0 to ø13.0 SCFP2013 ø12.0 to ø13.0
20
ø13.0 to ø14.0 SCFP2014 ø13.0 to ø14.0
25
ø14.0 to ø15.0 SCFP2015 ø14.0 to ø15.0
20
ø15.0 to ø16.0 SCFP2016 ø15.0 to ø16.0
20
*:
When a cable O.D. falls under two categories, try both and choose the more suitable one.
For Ethernet cable, use sealing fittings in the accessories kit for the connection port (see Figure
2.19). Remove the attached cable packing adapter and mount the sealing fitting back in place.
Six connection ports are provided in the electronics section. Use convenient ones.
Sep. 07, 2017-00