145
IM FG410-01EN
7.10 Sequence
Number of saved sequences
10 sequences (saved to non-volatile memory)
Maximum number of steps
Maximum of 255 steps per sequence (not including step of pre-
start status)
Inter-channel operation
In sequence mode, the mode of both channels is the sequence
mode. Step control is done in common for both channels.
Step control parameters
Step time, hold operation, jump destination, number of jumps, step
stop phase, branch operation, step termination control, step sync
code output
Intra-step channel parameters
Waveform, frequency, phase, amplitude, DC offset, square wave
duty
In-step operations
Constant, keep, linear interpolation (except waveform switching)
Step time setting range
0.1 ms to 1,000 s (4 digits or 0.01 ms resolution)
Jump count setting range
1 to 999 or infinite
Step stop phase setting range
0.000° to 360.000° (CH1 reference phase. 0.001° resolution) or
invalid
Branch operation
State branch
Check of state branch input from multi-I/O connector at step end.
Upon branch input detection, branching to specified destination
step
Event branch
Immediate branching to specified destination step through event
branch manipulation or input
Control of step termination
Sequence end or transition to next step
Step sync code output
Output of 4-bit code specified for each step to multi-I/O connector
LSB outputtable to sync/sub-output connector
Usable waveforms
Sine wave, square wave, noise, DC, and arbitrary wave
Ramp wave and parameter-variable waveform can be used through
saving as arbitrary waveforms.
Maximum number ofusable waveforms
128
Step start phase
Oscillation start from reference phase 0° of each channel at next
step after DC or noise (excluding DC and noise)
Sequence manipulations
Start, stop, hold/resume, event branch
Sequence external control
Input connector
Use of 4 bits of multi-I/O connector
Control items
Start or state branch, stop, hold/resume, event branch
Sequence external trigger input (start trigger)
Polarity
Positive, negative, off (selectable)
Input connector
Use of external trigger input on CH1 side. Input voltage and input
impedance follow the external trigger input specifications.
7.10 Sequence
Summary of Contents for FG420
Page 10: ...VIII IM FG410 01EN...
Page 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Page 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Page 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Page 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Page 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Page 173: ...159 IM FG410 01JA...
Page 178: ......
Page 179: ......