50
IM FG410-01EN
4. BASIC OPERATION
b) Items that can be changed in the phase setting
The following items can be changed with the phase setting.
The phase difference between the sync/sub-output’s
reference phase sync output
and waveform output can be changed
The following example assumes a phase setting of +90°. At this time, the waveform output’s zero
phase position is 90° ahead of the rise of the reference phase sync output.
Waveform output
Sync/sub-output
(reference phase sync output)
Phase setting: +90°
90°
Waveform output’s
zero phase position
The oscillation start/stop phase for burst oscillation, gated single-shot sweep can
be changed
The following example is burst oscillation with the oscillation start/stop phase set to
+30°. At this time, the oscillation starts at +30° and stops at the same +30° position.
Burst oscillation:
p.109 , gated single-shot sweep: p.93 .
Phase setting: +30°
Oscillation start at +30°
Oscillation stop at +30°
Summary of Contents for FG420
Page 10: ...VIII IM FG410 01EN...
Page 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Page 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Page 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Page 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Page 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Page 173: ...159 IM FG410 01JA...
Page 178: ......
Page 179: ......