93
IM FG410-01EN
Single-shot shuttle sweep
Sweep value
Frequency sweep example
Start value
Stop value
Sweep time
Sweep time
Trigger
j) To output the waveform only during sweep execution → Gated single-shot sweep
Set [SwpMode] (sweep mode) on the 2nd page of the setting screen to [Gated] (gated single-shot).
This operation combines gated oscillation and sweep, with sweep execution in synchronization with a
trigger.
Since a trigger signal is required, set the trigger signal with [Trig] (trigger) on the 2nd page. For the
trigger setting procedure,
Oscillation start/stop phase
The oscillation start/stop phase setting is done with [Phase] (phase) on the 1st page of the setting
screen.
However, during phase sweep, the start phase setting is the oscillation start phase, and the stop
phase setting is the oscillation stop phase.
Stop level (normally set to [Off])
To set the level while the oscillation is stopped separately from the phase, set [StpLvl] (stop
level) on the 3rd page of the setting screen to [On] (on) and set the level with a % value using the
amplitude full scale as reference. Normally, [Off] (off) is set. When [Off] is selected, the signal
level while oscillation is stopped is determined by the phase set in [Phase] on the 1st page of the
setting screen. For details on the stop level,
Oscillation stop unit (normally set to [Cycle] (1 cycle))
To stop oscillation in half-cycle units, set [OscStop] (oscillation stop unit) on the 3rd page of the
setting screen to [HalfCycle] (half cycle). Normally, set to [Cycle] (1 cycle). When [Cycle] (1
cycle) is set, whole integer cycle oscillation results.
When oscillation is stopped, it always stops either in 1-cycle or half-cycle units according to the
[OscStop] (oscillation stop unit) setting, and the oscillation time is usually longer than the sweep
time setting.
4.8 Sweep Setting and Manipulation
Summary of Contents for FG420
Page 10: ...VIII IM FG410 01EN...
Page 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Page 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Page 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Page 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Page 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Page 173: ...159 IM FG410 01JA...
Page 178: ......
Page 179: ......