IM 12D08N05-01E
2-2
<2. GENERAL SPECIFICATIONS>
Control function;
On/Off
Adjustable damping
Expire time
Hold;
Contact can be used to signal the Hold situation.
Fail safe;
Contact S4 is programmed as failsafe contact.
*1:
When contact output current is more than 4 Amps, ambient temperature should be less than
40 ºC.
Contact input:
Remote range switching to 10 times the programmed range.
Contact open;
If impedance > 100 kΩ: 1 x Range
(When programmed range for mA output is conductivity).
Contact closed;
If impedance < 10 Ω: 10 x Range
Temperature compensation:
Function;
Automatic or manual, for temperature ranges mentioned under B
(inputs).
Reference temp.; programmable from 0 to 100ºC or 30 to 210 °F (default 25ºC).
Compensation algorithm
:
According IEC 60746-3 NaCl tables (default). Two independent user
programmable temperature coefficients, from 0% to 3.5% per ºC (ºF) by
adjustment or calibration.
Matrix compensation:
With conductivity function of concentration and
temperature. Choice out of 13 preprogrammed
matrixes and 2 100-points user-programmable
matrixes.
Calibration
:
Semi-automatic calibration using pre-configured OIML (KCl) stanard tables,
with automatic stability check. Manual adjustment to grab sample.
Logbook:
Software record of important events and diagnostic data readily available in
the display or through HART
®
.
Display:
Graphical Quarter VGA (320 x 240 pixels) LCD with LED backlight and
touchscreen. Plain language messages in English, German, French,
Spanish, Italian, Swedish, Portuguese and Japanese.
Form:
Dimension;
144 (W) x 144 (H) x Approx.144 (D) mm
Converter weight; Approx. 1.5 kg
Housing:
Cast Aluminum housing with chemically resistant coating;
Polycarbonate cover with Polycarbonate flexible window
Protection IP66 / NEMA 4X / CSA Type 3S
Colour; Silver grey
SC450G-A(D)-A:
IP66 cable glands are supplied with the unit
SC450G-A(D)-U:
NEMA 4X close up plugs are mounted in the unused cable entry
holes and can be replaced by conduit fittings as required Pipe,
Panel or Wall mounting using optional hardware
Optional conduit adapter;
G1/2, 1/2NPT or M20 female
Power supply:
SC450G-A: