IM 12D08N05-01E
5-19
<5. MENU STRUCTURE COMMISSIONING>
l
HART
The address of the SC450G in a HART network can be set. Valid addresses are 0...15.
l
Factory adjustment
This menu is for service engineers only. This section is protected by a password.
Attempting to change data in the factory adjustment menu without the proper instructions
and equipment, can result in corruption of the instrument setup, and will impair the
performance of the unit.
5.15 Display setup
l
Main display
The main display consists of three lines with Process Values. Each line is user definable
with the restriction that each line should have a different Process Value. The default
settings can be defined here. By pressing one of the two smaller process values, this will
become the main process value in the main screen. Auto return will cause the main display
to go to default setting.
See also 4.6 Secondary to Primary Value display Switch.
NOTE
Configuration possibilities in the main and secondary display lines are determined by the
choices made in the menu measurement
Measurement setup >> Measurement
l
Additional text
Each process value can be given an additional text containing up to 12 characters per text.
This text is displayed on the main display next to the process value. This way the user can
distinguish separate measurements.
l
X-axis timing
The time range of the trend graph can be set from 15 minutes up to 14 days.
l
Y-axis limits
The ranges for each measurement need to be set according the application.
l
Auto return
When Auto return is enabled, the converter reverts to the measuring mode (main display)
from anywhere in the configuration menus, when no button is pressed during the set time
interval of 10 minutes. The HOLD flag will be cleared and all outputs will function normally.