IM 12D08N05-01E
APP-4
<APPENDIXES>
n
Temperature compensation matrix.
1. A minimum number of values is required to make interpolation possible. The
highlighted values marked as
are mandatory to enter.
Concent
Tref
T1
....
Tx
....
T10
Sol1
C1
S1Tr
S1T1
S1T10
....
Solx
SxTx
....
Sol10
C10
S10Tr
S10T1
S10T10
2. Tref (reference temperature) is defined in the Temperature Compensation menu. If
Tref is between T1 and T10 then the value of Tref needs to be entered as Tx (T2....T9).
Concent
Tref
T1
....
Tx
.... T10
Sol1
C1
S1Tr
S1T1
S1T10
....
Solx
....
Sol10
C10
S10Tr
S10T1
S10T10
3. For every SxTx that is entered the following values become mandatory to enter: Cx,
SxTr, SxT1, SxT10 and Tx.
Concent
Tref
T1
....
Tx
.... T10
Sol1
C1
S1Tr
S1T1
S1T10
....
Solx
Cx
SxTr
SxT1
SxTx
SxT10
....
Sol10
C10
S10Tr
S10T1
S10T10
The matrix can be cleared before
entering new values. Next new matrix
values can be entered as described
above. The SC450G can interpolate the
matrix. During this process it will check
if the matrix is completely ascending/
descending. This is necessary as
otherwise the lookup function can give
two results for one temperature. If an
error is found, the SC450G will specify
the location of the error as shown in the
user interface screen “user defined 1/2”.
The backspace key should be used for
deleting an individual matrix value.
An empty value is shown as
.