3.1.3 Installation
Installation of the Surface Scattering Light Turbidity Meter differs depending on whether the
sampling system is provided or not. If the sampling system is provided, secure the entire system
with anchor bolts (M12 (12 mm) bolts) to a well-drained “concrete base”. If the sampling system
is not provided, mount the detector and converter onto pipes (2-inch pipe) or racks with the
special mounting bracket. (See Model Code Option Specification.)
In the case of TB400G-
-
-KC (for Korea), a noise filter assembly is attached. Please refer to a
figure of external dimensions for the attachment means of the noise filter assembly.
50 mm dia. pipe
Pipe mounting bracket
Converter
Detector
Pipe mounting bracket
50 mm dia. pipe
F0301.ai
Figure 3.1 Pipe Mounting
Converter
Rack mounting bracket
M5 mounting screws (4 places)
238
126.5
4-M5
Mounting dimensions
Unit: mm
Detector
M5 mounting screws (3 places)
Rack mounting
bracket
190
105
3-M5
Mounting dimensions
F0302.ai
Figure 3.2 Rack Mounting