<2. SPECIFICATIONS>
2-2
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
Failure signal:
Transmitted continuously when failure occurs. Transmitted in
succession when multiple failures occur.
#ERR**
C
R
L
F
, error numbers (11, 12, 13, 14, 15, 18, 25, 26) in **.
Example 1: #ERR11
C
R
L
F
Example 2: #ERR11_ERR18
C
R
L
F
Note: A space code is transmitted by “_”
Data Update Interval: Approx. 1 second
Cable Length:
10 m max.
Contact Output:
Maintenance output (during maintenance)
Fail output (when an error is detected)
Range output (corresponding to the output range selected) (shared COM)
High/low limit alarm (default) or auto calibration/cleaning output,
selectable.
Type of contact output
Power off
Power on *1
Not in Action
In Action
Maintenance
Closed
Open
Closed
Fail
Open
Closed
Open
High/low alarm
Closed
Closed
Open
Auto zero calibration/cleaning
Closed
Open
Closed
*1:
Contact status (open/closed) when power is turned on is user configurable.
Contact Rating:
Maximum switching voltage: 250 V AC or 220 V DC (resistive load)
Maximum carrying current: 2 A AC or 2 A DC (resistive load)
Maximum switching capacity: 120 VA or 60 W (resistive load)
Contact Input:
Remote range switching (shared COM)
On input resistance: 200 Ω or less
Off input resistance: 100 kΩ or more
F0200.ai
R1
R2
R3
S1
S2
S3
Converter
S1: Output range 1
S2: Output range 2
S3: Output range 3
Remote Range Switching Diagram
Switch
Converter's Functions:
Display Function:
Data:
LED display – turbidity, lamp voltage, turbidity element detecting
current, reference element detecting current, analog output in %,
analog output in mA, zero point error and slope.
Operation status:
Lamp – <MEAS.>, <MAINT>, <PROGRAM 1> and <PROGRAM 2>
modes.
Indicator – output signal, HOLD, CAL, WASH, FAIL
Instrument status:
Indicator – lamp and each solenoid valve.
Maintenance Functions (<MAINT.> mode):
Zero calibration, shift calibration, span calibration, sensitivity correction
calibration, manual start of auto cleaning, manual start of auto zero
calibration, error code indication, error reset and calibration of lamp
control reference value.
Setting Functions:
<PROGRAM.1> mode: Setting of output range, average coefficient, output value in
failure condition, bi-linear output, high alarm, low alarm and spike
detection function.
<PROGRAM.2> mode: Setting of auto cleaning time and auto zero calibration time, and
selection of enabling/disabling failure detection functions.