<6. PROCEDURES FOR CONVERTER OPERATION>
6-31
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
(21) Setting Display-hold in Automatic Cleaning and Calibration
(FUNCTION "8.")
When "ON(execution)" is selected in this function, a turbidity value immediately before the
operation holds and displays during automatic cleaning and calibration in <MEAS.> mode.
The following shows the setting key operations.
(Example of display)
(1) Select FUNCTION "8." in the <PROGRAM1>
=> OFF (stop)
mode. The current set status is displayed.
(2) Press the [>] key to select on or off.
=> ON (execution)
F6601.ai
ON
OFF
(3) Press the [ENT] key to enter the setting.
=> ON
The display flashes once.
(22) Selecting Suffix Codes (sampling specifications) (FUNCTION "9.")
Select sampling specifications (Suffix codes). Usually, this is set upon shipment from the factory.
Thus, no setting is necessary before use. Table 6.11 shows the relationship between sampling
specifications and Suffix code.
Table 6.11 ME Codes and Settings
Suffix code
Sampling specification
Setting code
-NN, -KC
Without sampling system
_ _1
-A1
With sampling system and automatic cleaning
_ _1
-A2
With sampling system
_ _2
-A3
With sampling system and automatic cleaning and calibration
_ _3
If a change is necessary, perform the key operations in the following procedure.
(Example of display)
(1) Select FUNCTION "9." in the <PROGRAM1>
=> _ _3
mode. The current set code is displayed.
(2) Press the [>] key to select the setting code.=>
_ _2
F6604.ai
_ _1
_ _2
_ _3
(3) Press the [ENT] key to enter the setting.
=> _ _2
The display flashes once.