<2. SPECIFICATIONS>
2-3
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
Automatic Cleaning Function: Water jet cleaning. Cleaning time and interval are user
configurable. (When auto cleaning is specified)
Automatic Zero Calibration Function: Zero point calibration using zero water (When auto
cleaning/zero calibration is specified)
Error Detection Function: Turbidity overrange, lamp disconnection, lamp voltage failure, AD
circuit failure, memory failure and CPU failure.
Check Function:
Converter operation check
Manual Calibration:
Zero calibration:
Using zero water or light source off zero (selectable)
Span calibration:
Using calibration plate
Other Functions
Breakpoint line-segment (bi-linear) output, upper and lower limits alarms
Material:
Detector:
Modified black PPE (wetted part)
Piping:
Hard PVC, polyethylene resin, and polypropylene resin (all for
wetted parts)
Stanchion:
Carbon steel or stainless steel
Converter:
Aluminum alloy casting
Finish and Color:
Converter:
Baked polyurethane resin finish
Color; Munsell 0.6GY3.1/2.0 and Munsell 2.5Y8.4/1.2
Stanchion:
Baked polyurethane resin finish
Color; Munsell 0.6GY3.1/2.0
Ambient Temperature: -5 to 50°C (sample and tap water freeze in winter may need protection
against freezing)
Ambient Humidity: 5 to 95%RH (non-condensing)
Storage Temperature: -30 to 70°C
Installation Location: Indoors (For outdoor installation, a protection against rain is required
separately)
Installation:
Detector and converter: Pipe- or rack- mounted
With sampling system: Anchor bolt mounting
Piping Connection:
With sampling system: Drain: VP40
Others: VP16
Without sampling system: 25 mm ID hose joint
Cable Inlet:
5 cable glands (bottom of the converter)
Applicable wire:
6 to 12 mm, 9 to 11 mm when with arrester option
Wiring:
Power, analog output, digital output, contact output, contact input
and ground (using the ground terminal inside the converter or the
ground terminal on the outside of the case)
Power Supply:
100/110 V AC, 50/60 Hz, or 200/220 V AC, 50/60 Hz
Noise filter assembly: (only for TB400G-□-□-KC)
Ambient temperature: -10 to 50°C (no dew condensation allowed)
Storage temperature: -25 to 70°C
Construction:
JIS C 0920 Watertight (IP53)
Power Consumption:
Detector and converter: 50 VA or less, respectively
With sampling system: 200 VA or less (in full specifications, excluding combination
instruments)
Weight:
Detector: Approx. 3 kg
Converter: Approx. 9 kg
With sampling system: Approx. 50 kg (in full specifications, excluding
combination instruments)
Dimensions:
Detector: 245 (W) x 250 (H) x 200 (D) mm
Converter: 260 (W) x 340 (H) x 150 (D) mm