IM 01C22K01-01E
8-10
8. BRAIN TERMINAL BT200 OPERATION
(6) Integral Indicator Scale Setup
The following 5 displays are available for integral
indicators.
T0809.EPS
D20: DISP SELECT
and Display
NORMAL %
USER SET
USER & %
INP PRES
PRES & %
Indicates –5 to 110% range
depending on the Measurement
range (C21, C22).
Indicates values depending on the
Engineering range (D22, D23).
(
Note 1
)
Units set using Engineering unit
(D21) are not indicated.
Indicates user set and % alternately
in 3 second intervals.
Indicates input pressure.
(
Note 2
)
Indication limits –19999 to 19999.
Indicates input pressure and %
alternately in 3 second intervals.
A10:OUTPUT (%)
45.6 %
Description
and Related parameters
A11:ENGR.OUTPUT
20.0 M
A10:OUTPUT (%)
45.6 %
A11:ENGR. OUTPUT
20.0 M
A40:INPUT
456 kPa
A10:OUTPUT (%)
45.6 %
A40:INPUT
456 kPa
(Note 1) Scale range can be specified with range limit specifications up
to 5 digits (excluding any decimal point) for low or high range
limits within the range of –19999 to 19999.
The range with decimals is available to the third decimal place.
(Note 2) It indicates the value after zeroing.
See (a.) through (c.) for each setting procedure.
• Example: Change the low cut setting range
from 10% to 20%, and the low
cut mode from LINEAR to ZERO.
FEED
NO
OK
SET
D11:LOW CUT MODE
ZERO
Use the or key
to select “ZERO.”
Press the key twice to
enter the setting.
Press the (OK) key.
F0818.EPS
FEED
NO
OK
SET
D10:LOW CUT
20.0 %
ESC
SET
D11:LOW CUT MODE
LINEAR
< LINEAR >
< ZERO >
DATA
DIAG
PRNT
ESC
PARAM
D10:LOW CUT
20.0 %
D11:LOW CUT MODE
ZERO
D20:DISP SELECT
NORMAL %
CLR
ESC
SET
D10:LOW CUT
10.0 %
+ 20.0
Set “20.”
Press the key twice to
enter the setting.
Press the (OK) key.
Next, the [D11: LOW CUT MODE]
setting panel is displayed.
a. Display Selection (D20: DISP SELECT)
Follow the instructions given to the below to change
the range of integral indication scales.
When
USER SET
is selected, the user set values of
integral indication and
A11: ENGR. OUTPUT
parameter are indicated.
• Example: Set the integral indicator scale to
engineering units display.
The “%” disappears from the
integral indicator display.
ESC
SET
D20:DISP SELECT
USER SET
FEED
NO
OK
SET
D20:DISP SELECT
NORMAL %
<NORMAL %>
<USER SET>
<USER & %>
<INP PRES>
Use the or key
to select “USER SET.”
Press the key twice to
enter the setting.
Press the (OK) key.
F0820.EPS