<7. Operation>
7-7
IM 01C27C01-01EN
7.6.4 Venting Gas for Flushing Connection
Ring
1) Gradually open the vent screw to vent gas from
the flushing connection ring.
2) When the flushing connection ring is completely
vented, close the vent screw.
3) Tighten the vent screw to a torque of 10 N·m
{1 kgf·m}.
F0713.ai
Vent screw
When you loosen the vent screw,
the accumulated liquid(or drain)
will be expelled in the direction
of the arrow.
Figure 7.9
Venting for Flushing Connection Ring
WARNING
Since the accumulated liquid (or gas) may be
toxic or otherwise harmful, take appropriate care
to avoid contact with the body, or inhalation of
vapors.